COMPANY
TechPro
Operation
Manual
Model AR1
Zobrazit stránku 0
1 2 3 4 5 6 ... 11 12

Shrnutí obsahu

Strany 1 - Model AR1

10-02 (WEB-EG)MYRON LCOMPANYTechPro™OperationManualModel AR1

Strany 2 - MYRON L

change with concentration and temperature instead of single averagevalues.C. An Example of 2 different solution selections and the resulting compensat

Strany 3

their solution, like the KCl favored by chemists for its stability. Users dealing with sea water, etc., use NaCl as the model for theirconcentration c

Strany 4

XV. GLOSSARYAnions - Negatively charged ions. See Solution Characteristics, pg. 17.Algorithm - A procedure for solving a mathematical problem.See Te

Strany 5

Instrument Illustration11-12-99User selectable temperaturecompensationratios (SolutionSelection)Measurement key: User selectable for either Condu

Strany 6

FEATURES and SPECIFICATIONSA. Features• Superior resolution 3 1/2 digit LCD• Accuracy of ±1% of full scale• All electrodes are internal for maxi

Strany 7

TABLE OF CONTENTSInstrument Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1FEATURES and SPECIFI

Strany 8

I. INTRODUCTIONThank you for selecting the TechPro™ Series, Model AR1, one of theMyron L Company’s latest in a new line of digital instruments utilizi

Strany 9 - Temperature

V. SOLUTION SELECTIONA. Why Solution Selection is AvailableConductivity and TDS require temperature correction to 25°C values (ref.Standardized to 25°

Strany 10

C. Calibration Proceduresa. Rinse conductivity cell three times with proper standard (KCl, NaCl or 442) (ref. Conductivity/TDS Standard Solutions, pg.

Strany 11

A. Temperature ExtremesSolutions in excess of 160°F/71°C should not be placed in the cell cuparea; this may cause damage. Care should be exercised not

Strany 12

XI. ACCESSORIESA. Conductivity/TDS Standard SolutionsYour AR1 has been factory calibrated with the appropriate Myron LCompany NIST traceable standard